TROUBLESHOOTING GUIDE
VI .
TROUBLESHOOTING
VI.
1-5
Page
Phenomenon Cause Treatment
Manual Ref. Pages
17. FLUOVIEW software
cannot be started.
The FLUOVIEW software
has already been started.
Press Alt
+ Tab on
the keyboard to switch to
FLUOVIEW.
Another application is
running.
Exit from the running
software before starting the
FLUOVIEW software.
1-2-2, OPERATION
INSTRUCTIONS
18. The mouse pointer in the
screen cannot be
moved by moving the
mouse.
The software is
malfunctioning.
Press Ctrl
+ Alt +
Delete
on the keyboard
and exit from FLUOVIEW by
following the displayed
messages.
Then exit from windows and
reboot the computer.
1-2-1 & 1-2-2,
OPERATION
INSTRUCTIONS
19. The power supply of
Power Unit(FV5-PSU)
is not turned on.
The fuse has run out. Contact your local Olympus
representative for assistance
then replace the fuse.
-
20. Bios some check error
message appears at the
moment of starting up
the computer.
The battery maintains Bios
set value has run out.
Contact your local Olympus
representative for assistance
to replace the battery.
-
Bios has broken. Consult Olympus. -