APPLIED OPERATIONS/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2-89
Page
When click the mouse on the X or Y Axis on the graph window, the
[Editing] dialog box appears and the graph parameters or the graph
display method can be modified.
8. After image acquisition, select <Normal> under the [Surface XY] option button in
the [Mode] group box in the [Acquire] panel to return from the image acquisition
mode.
2-2-11 Image Acquisition on Desired Line (XZ, XT or XZT Observation)
When the specimen is in curved form, the image can be acquired by drawing a desired
line on it.
1. Acquire an image in the XY observation mode. For detailed operation, see section
2-2-1, “Image Acquisition with XY observation”.
2. Select [Line XT] option button in the [Mode] group box in the [Acquire] panel or
<Free> under the [Depth XZ] option button. A line indicating the scanning range
appears in the [Live] panel.
(<Free> under the [Line XT] option button is selected in the following Fig.)
A line can be specified for one display.
3. Move the curved line to the area you want to observe. To move the curve, place
the mouse pointer in the box and drag the mouse.
TIP
TIP