EasyManua.ls Logo

Olympus FLUOVIEW FV300 - Example of Assembling a Protocol; Example of Setting Procedure of FRAP Observation

Olympus FLUOVIEW FV300
564 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPLIED OPERATIONS/ Protocol processor
IV.
OPERATION INSTRUCTIONS
IV .
2-176
Page
2-4-8 Example of assembling a protocol
This section describes an example of assembling a protocol.
1 Example of Setting Procedure of FRAP Observation
An example of editing a protocol is described below assuming the FRAP observation.
Set Zoom
ratio to 2
An image
acquisition
Wait fo
r
5 sec.
Image acquisition:
3 frames
(Photo bleaching)
Protocol
Track 1 Track 2
Track 3
Set Zoom
ratio to 10
Set Zoom
ratio to 2
Image
ac
q
uisition
Image
saving
Repeat 30 times
Image
ac
q
uisition
Wait for
5 sec.
Initial image acquisition: Track 1 sets the zoom ratio to 2 x, and acquires an image.
Photobleaching: Track 2 sets the zoom ratio to 10 x, and acquires image 3 times to
make photobleach on the specimen.
Recovery measurement: Track 3 sets the zoom ratio to 2 x, and acquire image 3 times
with 5 seconds interval to observe the recovery. Then, stores the
image.
Execute a XY repeated scanning and adjust the set values for image
acquisition before executing the protocol.
1- Setting the observation condition
Before setting and starting the FRAP observation using the Protocol Processor, execute
a repeated scan and set the condition for the observation.
Assuming FITC to Channel 2 and image size to 320X240 pixels, the following figure
shows an example of image acquisition.
TIP

Table of Contents