EasyManua.ls Logo

Olympus FLUOVIEW FV300 - Input Supporting Function

Olympus FLUOVIEW FV300
564 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPLIED OPERATIONS/Protocol processor
IV.
OPERATION INSTRUCTIONS
IV .
2-165
Page
One Point!
It is possible to use variables in [For] statement in other cell of protocol processor.
<E.g>
It is possible to change laser intensity value every image acquiring time. When laser intensity of argon laser is
defined as variable N, it creates protocols that acquire image with interval of step 10 by N equal to 10 to 100.
㪈㪅 Enter [N = 10 To 100 Step 10] for Condition column in No.1 track.
㪉㪅 Define laser intensity of argon laser as variable N for No.2 track.
㪊㪅 Define return of repetition for No.3 track.
㪋㪅 Select <Start> button.
Image will be acquired by changing laser intensity of argon laser by 10% as shown below.
1st 2nd 3rd 4th 5th 㵺㩷 10th
㪣㪸㫊㪼㫉㩷㩿㩼㪀㩷 㪈㪇㩷 㪉㪇 㪊㪇㩷 㪋㪇㩷 㪌㪇㩷 㵺㩷 㪈㪇㪇㩷
4 Input supporting function
[Track] window is available as input supporting function for each track.
[Track] window appears by double-clicking No of [PAPP] window.

Table of Contents

Other manuals for Olympus FLUOVIEW FV300

Related product manuals