EasyManua.ls Logo

Olympus FLUOVIEW FV300 - Differences in Image Acquisition Method between Fluorescent and Transmitted Images; Monochrome Image

Olympus FLUOVIEW FV300
564 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPLIED OPERATIONS/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2-56
Page
2-2-3 Differences in Image Acquisition Method Between Fluorescent and
Transmitted Images
2-2-3-1 Monochrome Image
The wavelength obtained by monochrome dyeing can be acquired and observed as an
image of a channel (either Ch1 or Ch2).
The subsequent description deals with the differences in operation between the
monochrome dyeing and dual-fluorochrome dyeing.
When the specimen is dyed so that the wavelength of the fluorescence emission from it
is below 570 nm, the image is acquired in channel 1 (Ch1).
When the specimen is dyed so that the wavelength of fluorescence emission from it is
above 570 nm, the image is acquired in channel 2 (Ch2).
z Set the [DETECTION MODE] lever of the scan unit to “FL -” or “- FL”.
FL -: When acquiring the image in Ch1.
- FL: When acquiring the image in Ch2.
z Engage the filter by setting the [BARRIER FILTERS] lever of the scan unit. (Setting
the lever to “IN” engages the filter.)
CH1: When observing a specimen dyed to generate fluorescence wavelength
below 570 nm.
CH2: When observing a specimen dyed to generate fluorescence wavelength
above 570 nm.
Also see section 2-2-1-2, “Configuring the Scan Unit“ and follow the instructions given
by the [Microscope Configuration] window.

Table of Contents

Other manuals for Olympus FLUOVIEW FV300

Related product manuals