Getting Started FLUOVIEW/Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV
. 1-30
Page
4. Focus on the specimen by looking into the eyepiece.
Be sure to adjust the diopter of the eyepiece in advance. (Refer to the
instruction manual of the IX70/81 microscope.)
When the Manual microscope and the Z motor are in use, clear
the check mark in the [Locked] check box in the [Z Stage]
sub-panel in the [Acquire] panel (see section 2-2-1-3-7 of this
volume), then focus on the specimen by operating the fine
focus adjustment knob of the microscope. Be sure not to
operate the fine focus adjustment knob while the [Locked]
check box is checked, for this may damage the Z motor.
The specimen may float during oil-immersed observation. In
this case, prepare a stage clip (U-SCL) and attach it to the
microscope as shown on the left.
NOTE