APPLIED OPERATIONS/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2-47
Page
1 Setting the observation mode
1. Click the [Surface] option button in the [Mode] group box in the [Acquire] panel.
2. In the [Acquire] panel, select the XYT observation mode option button.
2 Setting the interval time
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Time
Series] sub-panel. A panel as shown below will be displayed.
Fig. 2-15 [Time Series] Sub-panel
1. Set the interval time using the <
> or <
> button in the [Interval] text box.
An image with no interval can be acquired by entering [0] in the [Interval]
text box.
In this case, the [Interval] text box shows “Free Run” message.
Using <
> button also sets the [Interval] text box to “Free Run”.
TIP
[Rest Time] text box
Shows the time after end of acquisition
of an image until the start of next
acquisition.
[N] text box
Set the number of scans using the <
>
or <
> button or by input from the
keyboard.
Under text box, it displays max. number
of slices that can be acquired with use
of physical memory only available at
that time.
[Interval] text box
Set the interval time using the <
> or
<
> button or by input from the keyboard.
[Total Time] text box
Shows the total time required for image
acquisitions.