APPLIED OPERATIONS/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2-106
Page
6. Start the third scan in 10 seconds after the second scan is performed.
(The [Disabled] option button is selected.)
7. Start the forth and subsequent scan in 10 seconds after the third scan is
performed.
The image of the photobleached specimen is acquired.
When performing the forth or subsequent scan, be sure to switch the
[REX mode] and [Bleach mode] between scans.
In the TIME COURSE software (optional), when the REX mask file in
the REX or Bleach mode is changed while image scanning is
performed obtaining the real time graph, the laser is irradiated while
the pop-up menu is displayed, however, the real time graph is not
plotted. Please be aware this before using.
The image of the specimen
is fading only where the
laser is irradiated.
NOTE
NOTE