Getting Started FLUOVIEW/Basic Operations
IV.
OPERATION INSTRUCTIONS
IV .
1-5
Page
(3) Barrier filter setting
The barrier filters can be switched over by pushing or pulling the BARRIER
FILTERS slider (3).
Up to 2 barrier filters can be mounted for each of CH1 and CH2. The filters are
engaged in the light path when the slider switch is pushed in and disengaged when
it is pulled out.
When combination is not known, see section 1-3-2-3, "Configuring the Filters" in
this volume and follow instructions in the [Optical System Configuration] window.
If the slider switch is set pushed-in harf way to the position by
mistake, no images can be observed at all. Be sure to set it to
the fully pulled-out or pushed-in position
(4) Switching the excitation DM
One of the two DMs can be selected using the excitation DM selector knob f. As
the light axis may be deviated when the excitation DM is switched with the
excitation DM selector knob, adjustment of the detection light axis is required after
switching.
For details on detection light axis adjustment, see section 1-2-12, “Adjustment of
Detection Light Axis” in the Main Volume.
Overlapping the image before excitation DM switching and that
after it may result in deviation between images. To prevent
this, it is recommended to use excitation DMs that can be
excited simultaneously
.
NOTE
NOTE