Getting Started FLUOVIEW/Basic Operations
IV.
OPERATION INSTRUCTIONS
IV .
1-16
Page
(5) Condenser, polarizing plate
Used for transmitted lighting.
In addition, the rotary turret for the transmitted light DIC slider and the polarizing
plate for differential interference observation (analyzer) are also provided.
x To perform differential interference observation, engage the transmitted light
DIC slider (optional) matching the objective in use in the light path. (For both
direct observation and laser differential interference observation)
x To perform direct differential interference observation or laser differential
interference observation, engage the polarizing plate in the light path.
(6) Filters
These filters are used to adjust the transmitted light.
x For transmitted observation using laser, disengage the LBD filter from the light
path and engage the FR (frost) filter in the light path by operating the filter
levers. If the FR filter is disengaged from the light path, the image may be
marred by stripe interference.
(7) Hand switch U-HSTR2 (U-FH is optionally available.)
This is the hand switch to operate the IX motorized system.