EasyManua.ls Logo

R&S FSV3000 Series - Page 1047

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1047User Manual 1178.8520.02 ─ 01
CALCulate<n>:LIMit<li>:LOWer:MODE <Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST: ABSolute
Manual operation: See " X-Axis " on page 479
CALCulate<n>:LIMit<li>:LOWer:OFFSet
<Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
Manual operation: See " Y-Offset " on page 477
CALCulate<n>:LIMit<li>:LOWer:SHIFt <Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
Window
<li> Limit line
Analyzing Measurements (Basics)

Table of Contents

Other manuals for R&S FSV3000 Series

Related product manuals