Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
479User Manual 1178.8520.02 ─ 01
Remote command:
CALCulate<n>:LIMit<li>:LOWer:THReshold on page 1048
CALCulate<n>:LIMit<li>:UPPer:THReshold on page 1052
Margin
Defines a margin for the limit line. The default setting is 0 dB (i.e. no margin).
For details on margins see "Limits and Margins" on page 471.
Remote command:
CALCulate<n>:LIMit<li>:LOWer:MARGin on page 1046
CALCulate<n>:LIMit<li>:UPPer:MARGin on page 1050
X-Axis
Describes the horizontal axis on which the data points of the limit line are defined.
Includes the following settings:
Unit:
"Hz" : for frequency domain
"s" : for time domain
Scaling mode: absolute or relative values
For relative values, the frequencies are referred to the currently set center fre-
quency. In the time domain, the left boundary of the diagram is used as the refer-
ence.
Scaling: linear or logarithmic
Remote command:
CALCulate<n>:LIMit<li>:LOWer:MODE on page 1047
CALCulate<n>:LIMit<li>:UPPer:MODE on page 1050
CALCulate<n>:LIMit<li>:CONTrol:DOMain on page 1044
CALCulate<n>:LIMit<li>:CONTrol:SPACing on page 1045
Y-Axis
Describes the vertical axis on which the data points of the limit line are defined.
Includes the following settings:
Level unit
Scaling mode: absolute or relative (dB/%) values
Relative limit values refer to the reference level.
Limit type: upper or lower limit; values must stay above the lower limit and below
the upper limit to pass the limit check
Remote command:
CALCulate<n>:LIMit<li>:UNIT on page 1049
CALCulate<n>:LIMit<li>:LOWer:SPACing on page 1048
CALCulate<n>:LIMit<li>:UPPer:SPACing on page 1051
Data Points
Each limit line is defined by a minimum of 2 and a maximum of 200 data points. Each
data point is defined by its position (x-axis) and value (y-value). Data points must be
defined in ascending order. The same position can have two different values.