Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1049User Manual 1178.8520.02 ─ 01
<li> Limit line
Parameters:
<Threshold> Numeric value.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 1049.
*RST: -200 dBm
Default unit: DBM
Manual operation: See " Threshold " on page 478
CALCulate<n>:LIMit<li>:NAME
<Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Name> String containing the limit line name.
*RST: REM1 to REM8 for lines 1 to 8
Manual operation: See " Name " on page 478
CALCulate<n>:LIMit<li>:UNIT <Unit>
This command defines the unit of a limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Unit> DBM | DBPW | WATT | DBUV | DBMV | VOLT | DBUA |
AMPere | DB | DBUV_M | DBUA_M | (unitless)
If you select dB as the limit line unit, the command automatically
turns the limit line into a relative limit line.
*RST: DBM
Manual operation: See " Y-Axis " on page 479
CALCulate<n>:LIMit<li>:UPPer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Analyzing Measurements (Basics)

Table of Contents