Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1048User Manual 1178.8520.02 ─ 01
Parameters:
<Distance> Defines the distance that the limit line moves.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 1049.
Default unit: DB
Manual operation: See " Shift y " on page 480
CALCulate<n>:LIMit<li>:LOWer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of a lower
limit line from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See " Y-Axis " on page 479
CALCulate<n>:LIMit<li>:LOWer:STATe <State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<li>:NAME on page 1049.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See " Visibility " on page 476
CALCulate<n>:LIMit<li>:LOWer:THReshold <Threshold>
This command defines a threshold for relative limit lines.
The R&S FSV/A uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant