Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1102User Manual 1178.8520.02 ─ 01
For details, see Chapter 12.6, "Frequency Response Correction (R&S FSV/A-K544)",
on page 604.
● Remote Commands for Frequency Response Correction.................................. 1102
● Programming Example: Using Touchstone files.................................................. 1116
14.10.4.1 Remote Commands for Frequency Response Correction
The following commands are only available if the Frequency Response Correction
option (R&S FSV/A-K544) is installed.
Input-specific frequency correction
Frequency response correction can be configured for all inputs, or for particular input
types only. Be sure to use the correct command for the required setting.
●
[SENSe:]CORRection:FRESponse:INPut<ip>:USER: commands are applied
to RF input only
●
[SENSe:]CORRection:FRESponse:BASeband:USER: commands are applied
to baseband input only
●
[SENSe:]CORRection:FRESponse:USER: commands are applied to all input
types; queries refer to the currently active input source
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:CATalog?.........................1104
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:CATalog?.......................... 1104
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:CATalog?......................................... 1104
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:CLEar..............................1104
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:CLEar...............................1104
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:CLEar..............................................1104
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:DATA:FREQuency?...........................1104
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:DATA:MAGNitude?........................... 1104
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:DATA:PHASe?..................................1104
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:INSert..............................1105
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:INSert...............................1105
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:INSert..............................................1105
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:MAGNitude[:STATe]..........1105
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:MAGNitude[:STATe]........... 1105
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:MAGNitude[:STATe].......................... 1105
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:PHASe[:STATe]................ 1106
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:PHASe[:STATe]................. 1106
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:PHASe[:STATe].................................1106
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fi>:REMove...........................1106
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:REMove........................... 1106
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:REMove...........................................1106
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:SELect............................ 1106
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:SELect............................. 1106
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:SELect.............................................1106
[SENSe:]CORRection:FRESponse:BASeband:USER:FLISt<fli>:SIZE?..............................1107
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:FLISt<fli>:SIZE?...............................1107
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:SIZE?.............................................. 1107
Configuring the R&S
FSV/A