EasyManua.ls Logo

R&S FSV3000 Series - Page 1148

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1148User Manual 1178.8520.02 ─ 01
INSTrument:COUPle:USER<uc>:NUMBers:LIST?
This command queries the index numbers of user-defined parameter couplings. The
index numbers are used to refer to the specific coupling in remote commands with a
USER<uc> suffix.
Suffix:
<uc>
.
irrelevant
Return values:
<Index> Comma-separated list of strings
Index numbers of all available user-defined couplings
Note that the returned index numbers are not necessarily the
same as those shown in the user interface.
Example:
INST:COUP:USER:NUMB:LIST?
Result:
'1','2','4'
Number '3' is not returned, because a coupling with that index
does not exist anymore.
Usage: Query only
Manual operation: See "Index" on page 635
INSTrument:COUPle:USER<uc>:RELation <Direction>
This command selects the direction in which synchronization works.
Note that the command is not available if you synchronize over all channels or all
channels of the same application.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by INSTrument:COUPle:USER<uc>:
NUMBers:LIST?.
Parameters:
<Direction> LTOR | RTOL | BIDir
BIDir
Changes of a parameter are applied both ways (from channel 1
to channel 2 and vice versa).
LTOR
Changes of a parameter are applied from channel 1 to channel
2, but not the other way around.
RTOL
Changes of a parameter are applied from channel 2 to channel
1, but not the other way around.
Example:
INST:COUP:USER:REL BID
Selects bidirectional changes for the user-defined coupling with
the index number 1.
Manual operation: See "Direction" on page 635