Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
983User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>[:STATe]............................................................................... 987
CALCulate<n>:MARKer<m>:TRACe................................................................................987
CALCulate<n>:MARKer<m>:X........................................................................................ 987
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off all delta markers.
Suffix:
<n>
.
Window
<m> irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
CALCulate<n>:DELTamarker<m>:LINK
<State>
This command links delta marker <m> to marker 1.
If you change the horizontal position (x-value) of marker 1, delta marker <m> changes
its horizontal position to the same value.
Tip: to link any marker to a different marker than marker 1, use the CALCulate<n>:
DELTamarker<ms>:LINK:TO:MARKer<md> or CALCulate<n>:MARKer<ms>:
LINK:TO:MARKer<md> commands.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT2:LINK ON
Manual operation: See " Linking to Another Marker " on page 431
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md> <State>
This command links delta marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, delta marker <m1> changes its
horizontal position to the same value.
Suffix:
<n>
.
Window
<ms> source marker, see Marker
Analyzing Measurements (Basics)