EasyManua.ls Logo

Rohde & Schwarz SMBV100A - General Configuration of Instrument - Setup Key

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
105Operating Manual 1407.6062.32 ─ 08
5.2.3 General Configuration of Instrument - Setup Key
The SETUP key opens the "Setup" dialog.
The "Setup" dialog is divided into several sections.
The "System" section is used to set general instrument parameters.
The "Test" section is used to perform function tests.
The "Environment" section is used to configure the controller interfaces.
The "Remote" section is used to configure the remote control interfaces.
The "Protection" is used to set the protection level for service functions and security
settings.
The "Settings" section is used to save or recall instrument settings or to preset the
instrument to factory settings.
Most submenus of this key can be accessed only via the SETUP key or the menu tree
( MENU key), with the following exceptions:
The "Reference Oscillator" dialog can also be called up in the "RF" block and is
therefore described in the section on this block (see chapter 5.4.4, "Reference Oscil-
lator", on page 159).
The "Save/Recall" dialog can also be called up with the FILE key and is therefore
described in the section on this key (see chapter 5.2.8, "Storing and Loading Instru-
ment Data - File Key", on page 138.
The "Global Trigger/Clock/ External Inputs" dialog is also available in all modulation
menus of the "Baseband" function block where it can be called up with the "Global
Trigger/Clock Settings" button in the corresponding "Trigger/Marker/Clock" dialog.
See also chapter 5.2.3.12, "Global Trigger/Clock/External Input Set-
tings", on page 121.
5.2.3.1 Internal Adjustments
The R&S Signal Generator is extremely accurate thanks to the integrated procedures for
adjustments.
All internal adjustments for which no external measuring equipment is needed can be
started in the "Internal Adjustments..." dialog. The adjustments with external measuring
equipment are described in the Service Manual (on CD ROM, supplied with the instru-
ment).
Adjustment is recommended if the temperature range in which the instrument is operated
changes, or prior to all applications which require maximum level and frequency accu-
racy.
During adjustment a bar indicates the status of progress. If an error occurs, adjustment
is terminated and an error message is output in the info line.
General Instrument Settings

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals