Remote Control Commands
R&S
®
SMBV100A
545Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:BB:ARB:WSEGment:CONFigure:SEGMent:APPend <Waveform>
This command appends the specified waveform to the configuration file.
Setting parameters:
<Waveform> string
Example:
MMEM:CDIR '/var/smbv/user/waveform'
sets the default directory to /var/smbv/user/waveform.
BB:ARB:WSEG:CONF:SEL 'new'
creates the configuration file new.inf_mswv in the default direc-
tory.
BB:ARB:WSEG:CONF:SEGM:APP 'arb1'
appends waveform arb1.wv to configuration file new. Waveform
arb1 will be the first segment of a multi segment waveform cre-
ated with configuration file new.
Usage: Setting only
[:SOURce<hw>]:BB:ARB:WSEGment:CONFigure:SEGMent:CATalog?
This command queries the segments of the currently selected configuration file.
Return values:
<Catalog> string
Example:
MMEM:CDIR '/var/smbv/user/waveform'
sets the default directory to /var/smbv/user/waveform.
BB:ARB:WSEG:CONF:SEL 'multi_sin'
selects the configuration file multi_sin.inf_mswv .
BB:ARB:WSEG:CONF:SEGM:CAT?
queries the segments of the selected configuration file.
Response: arb4, arb2
The configuration file includes the segments arb4.wv and
arb2.wv.
Usage: Query only
[:SOURce<hw>]:BB:ARB:WSEGment:CONFigure:SELect <Filename>
The command selects the configuration file in the default directory A path can also be
specified, in which case the files in the specified directory are selected. If a configuration
file with the specified name does not yet exist, it is created. The file extension may be
omitted. Only files with the file extension *.inf_mswv will be selected or created.
Parameters:
<Filename> string
Example:
MMEM:CDIR '/var/smbv/user/waveform'
sets the default directory to /var/smbv/user/waveform.
BB:ARB:WSEG:CONF:SEL 'new'
create configuration file new.inf_mswv in the default directory.
SOURce Subsystem