Remote Control Commands
R&S
®
SMBV100A
719Operating Manual 1407.6062.32 ─ 08
:TEST<hw>:ALL:RESult?
Queries the result of the performed selftest (command :TEST<hw>:ALL:STARt
on page 718).
Return values: 
<All> 0 | 1
Example: 
TEST:ALL
Starts the selftest
:TEST:ALL:RES?
Response: "0" on success, "1" on fail
Usage:  Query only
:TEST<hw>:DIRect <HW_assembly>,<subadress>,<hex data string>
:TEST<hw>:DIRect? <HW_assembly>,<subadress>
The respective hardware assembly responds directly to the command; any safety mech-
anisms are bypassed. This function is only available via remote control.
Example: 
TEST:DIR 'SSYN',0,#H12345678
TEST:DIR? 'SSYN',0
Response: #H12345678
7.18 TRIGger Subsystem
The TRIGger system contains the commands for selecting the trigger source for the RF
and LF sweep. The trigger input connectors are configured in the SOURce:INPut sub-
system.
The trigger system of the R&S SMBV is a simplified implementation of the SCPI trigger
system. The TRIGger system differs from the SCPI system as follows:
●
No INITiate command; the instrument behaves as if INITiate:CONTinuous
ON were set.
●
Under TRIGger several sweep subsystems exist.
Other commands associated with the trigger system of the R&S SMBV can be found in
the modulation and RF signal subsystems.
TRIGger<hw>
●
Suffix TRIGger<1|2> is not permitted
●
TRIGger0 activates the LF output.
Table 7-3: Cross-reference between the manual and remote control
R&S name SCPI name Command under manual control
AUTO IMMediate "Auto" mode
SINGle BUS "Single" mode.
 
TRIGger Subsystem