Instrument Function
R&S
®
SMBV100A
267Operating Manual 1407.6062.32 ─ 08
Sets the output voltage for both signal components. Small differences at the inputs can
be taken into account via impairment settings (see chapter 5.6.1.1, "Impairment Settings
Menu", on page 249).
Single ended Output:
Value range: 1 mV ... 1.5 V EMF.
Differential Output:
Value range: 2 mV ... 3 V EMF.
SCPI command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP on page 641
Digital Attenuation
The level of the baseband and the RF signal is attenuated by this value.
SCPI command:
[:SOURce<hw>]:IQ:OUTPut:POWer:DATTenuation on page 642
Resulting I/Q Level Vp EMF
Displays the resulting output off-load voltage of both signal components at the rear panel.
SCPI command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP:RESult on page 641
5.7 Digital I/Q Output Settings
The output of the digital baseband signal requires the option R&S SMBV-K18, Digital
Baseband Connectivity. If this option is connected, the instrument adds as separate
function block to the block diagram, the "BB In/Out" block.
If digital Output is selected, the R&S SMBV displays "DIG I/Q OUT" and a corresponding
symbol in the "BB In/Out" block.
The digital baseband signal is output at a standardized (LVDS) digital I/Q interface.
The settings dialogs for the digital signal output is available in the "BB In/Out" function
block as well as in the menu with the same name which is opened using the MENU key.
To access the settings for the digital I/Q output:
1. select "Block Diagram > BB In/Out > Config... > Baseband Output Settings" or
2. press the MENU key and select "BB In/Out > Baseband Output Settings".
Digital I/Q Output Settings