[:SOURce]:INPut:DELay:CLENgth<ch> <CableLength>
Sets the total length of the cable used to connect the CLK OUT and CLK IN connectors.
For the first slave in the chain, the value represents the cable length between the master
and the first slave; for the second slave, the cable length to the first and to the second
has to be defined, etc.
Parameters:
<CableLength> float
Range: 0 to 10
Increment: 0.01
*RST: 0.5
Example: Setting in the first slave instrument Slave#1:
:SOURce:INPut:DELay:SPOSition 1
:SOURce:INPut:DELay:CLENgth1 1
:SOURce:INPut:DELay:ADCState ON
Example: Setting in the second slave instrument Slave#2:
:SOURce:INPut:DELay:SPOSition 2
:SOURce:INPut:DELay:CLENgth1 1
:SOURce:INPut:DELay:CLENgth2 0.5
:SOURce:INPut:DELay:ADCState ON
7.14.9 SOURce:IQ Subsystem
This subsystem contains the commands for checking the I/Q modulation.
[:SOURce<hw>]:IQ:CREStfactor......................................................................................637
[:SOURce<hw>]:IQ:IMPairment:IQRatio[:MAGNitude]........................................................638
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I.......................................................................638
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q.....................................................................638
[:SOURce<hw>]:IQ:IMPairment:QUADrature:[ANGLe].......................................................638
[:SOURce<hw>]:IQ:IMPairment[:STATe]..........................................................................639
[:SOURce<hw>]:IQ:SOURce...........................................................................................639
[:SOURce<hw>]:IQ:STATe..............................................................................................639
[:SOURce<hw>]:IQ:SWAP[:STATe].................................................................................640
[:SOURce<hw>]:IQ:WBSTate..........................................................................................640
[:SOURce<hw>]:IQ:CREStfactor <CrestFactor>
This command specifies the crest factor of the external analog signal.
Parameters:
<CrestFactor> float
Range: 0 dB to 35 dB
Increment: 0.01dB
*RST: 0 dB
Default unit: dB