EasyManua.ls Logo

Rohde & Schwarz SMBV100A - I;Q Settings

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
237Operating Manual 1407.6062.32 ─ 08
Use the "I/Q Mod In" section to select the input signal:
select "Analog Wideband I/Q In", if the external analog signal is fed into the I/Q mod-
ulator.
The maximum available RF bandwidth is described in the instruments specifications
(see data sheet).
select "Internal Baseband I/Q In", if the internally generated baseband signal is used.
The RF bandwidth is reduced (see data sheet).
Alternatively, the input signal can be selected in the "I/Q Settings" menu (see chapter 5.5,
"I/Q Modulation - I/Q Mod Block", on page 235).
5.5.2 I/Q Settings
To open the menu for setting the I/Q modulation parameters, select "I/Q Mod > Configure
> I/Q Settings" or use the MENU key.
System error correction of the I/Q modulator permits precise and repeatable measure-
ments. The correction routine should be called in the case of temperature fluctuations of
several degrees. To call the routine, press the SETUP key and select "System > Internal
Adjustment " and execute the desired adjustment procedure.
It is not possible to activate an I/Q modulation with an external analog signal ("Analog
Wideband I/Q In") simultaneously with:
AM,
BB-AM,
digital modulation,
digital standards,
noise,
arbitrary waveform generation and
multi carrier CW
I/Q modulation is activated and the source entered in the top section of the menu.
I/Q impairment for specific impairment of the I/Q modulation is set in the "Impairments"
section.
I/Q Modulation - I/Q Mod Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals