Remote Control Commands
R&S
®
SMBV100A
456Operating Manual 1407.6062.32 ─ 08
If you want to write an ASCII C-List directly without using the C-List editor, a special syntax
has to be taken into account:
The tag and the following [state: start time]-pairs are separated by a colon ":".
The [state: start time] pairs in turn are separated by semicolons ";".
The list state can be either 0 or 1.
The time unit is the simple period.
Example:
"CW MODE LIST 0:0;380:1" for a CW trace:
The list status starts at time 0 with 0, and at time point 380 symbol periods after start, it
changes to 1.
With the commands BB:DM:CLISt:... you can also generate control lists in binary
fomat, if required. See "DM Lists" on page 572.
7.4 CALibration Subsystem
The CALibration system contains the commands for adjustment. Adjustment is triggered
by the query commands. The response "0" indicates error-free adjustment, and the
response "1" means that an error occurred during adjustment.
:CALibration:ALL[:MEASure]...........................................................................................456
:CALibration<hw>:FMOFfset[:MEASure]...........................................................................457
:CALibration<hw>:FREQuency[:MEASure].......................................................................457
:CALibration<hw>:IQModulator:FULL...............................................................................457
:CALibration<hw>:IQModulator:LOCal..............................................................................457
:CALibration<hw>:LEVel:ALCTable[:MEASure].................................................................458
:CALibration<hw>:LEVel[:MEASure]................................................................................458
:CALibration<hw>:LEVel:EXTern:DATA............................................................................458
:CALibration<hw>:LEVel:STATe......................................................................................459
:CALibration:ALL[:MEASure]?
The command starts all internal adjustments for which no external measuring equipment
is needed.
Return values:
<Measure> 0 | 1
Example:
CAL:ALL:MEAS?
starts the adjustment of all functions for the entire instrument.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
SCPI confirmed
CALibration Subsystem