Preparing for Use
R&S
®
SMBV100A
35Operating Manual 1407.6062.32 ─ 08
"Modulations State" = Off
Uninterrupted level settings are switched off
"Level Attenuator Mode" = AUTO
Internal level control "Level ALC" = AUTO
User correction "Level Ucor" = OFF
"LF output State" = Off
"Sweep State" = Off
"List mode State" = Off
Settings that are not affected by the PRESET key
Reference frequency settings ("Ref Oscillator" menu)
Power on settings ("Level/EMF" menu)
Network settings ("Setup" menu)
GPIB address ("Setup" menu)
*IDN? Identification and emulation ("Setup" menu)
Password and settings protected by passwords ("Setup" menu)
Start/Stop Gui Update ("Setup" menu)
Display and keyboard settings ("Setup" menu)
User-defined instrument states can be stored and called up in the "File" dialog.
2.3.5 Shutting Down the Instrument
To shut down the R&S SMBV, proceed as described below.
Risk of losing data
If you switch off the running instrument using the rear panel switch or by disconnecting
the power cord, the instrument loses its current settings. Furthermore, program data may
be lost.
Press the ON/STANDBY key first to shut down the application properly.
1. Press the ON/STANDBY key to save the current setup, shut down the operating
system and set the instrument to standby state.
The yellow LED must be on.
2. To turn the power off:
a) For instruments with power switch, press the main power switch to position 0
(Off).
Putting into Operation