Instrument Function
R&S
®
SMBV100A
376Operating Manual 1407.6062.32 ─ 08
The Multi Carrier CW signal is only computed when the "Accept" button is pressed.
Whenever the table contains settings that have not yet been adopted with the "Accept"
button, the background is yellow.
Note: The phase/deg settings are only valid if optimization of the crest factor is disabled
("Optimize Crest Factor" = Off).
"No."
This is the carrier index.
"State"
Switch a carrier on/off
"Power"
Sets the power of a carrier.
"Phase"
Sets the starting phase of a carrier.
"Accept"
Transfer the settings in the carrier table into the instrument.
SCPI command:
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe on page 592
[:SOURce<hw>]:BB:MCCW:CARRier:STATe on page 594
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer on page 592
[:SOURce<hw>]:BB:MCCW:CARRier:POWer on page 593
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:PHASe on page 591
[:SOURce<hw>]:BB:MCCW:CARRier:PHASe on page 593
Carrier Graph
Calls a graphical representation of the chosen carrier configuration.
The carriers are on the X-axis and the colored bars represent those carriers which are in
the On state. Power is on the Y-axis, and the height of the bars corresponds to the chosen
power of each individual carrier.
SCPI command:
n.a.
Baseband Signal - Baseband Block