Remote Control Commands
R&S
®
SMBV100A
459Operating Manual 1407.6062.32 ─ 08
:CALibration<hw>:LEVel:STATe <State>
The command switches on or off internal level correction.
Parameters:
<State> ON | OFF
*RST: ON
Example:
CAL:LEV:STAT
switches on level correction.
7.5 CLOCk Subsystem
The CLOCk system contains the commands for configuration of the signals at the clock
output and input connectors.
:CLOCk:INPut:FREQuency.............................................................................................459
:CLOCk:INPut:SLOPe....................................................................................................459
:CLOCk:INPut:FREQuency?
The command sets the measured frequency of the external clock signal. An external clock
reference must be supplied at the CLOCK input.
Return values:
<Frequency> float
Example:
CLOC:INP:FREQ
queries the measured frequency of the external clock reference.
Usage: Query only
:CLOCk:INPut:SLOPe <Slope>
The command sets the active slope of an externally applied clock signal at the CLOCK
connector.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example:
CLOC:INP:SLOP NEG
The active slope of the external clock signal at the CLOCK con-
nector is the falling slope.
7.6 DIAGnostic Subsystem
The DIAGnostic system contains the commands used for instrument diagnosis and
servicing. SCPI does not define any DIAGnostic commands; the commands listed here
CLOCk Subsystem