Remote Control Commands
R&S
®
SMBV100A
457Operating Manual 1407.6062.32 ─ 08
:CALibration<hw>:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
Return values:
<Measure> 0 | 1
Example:
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Usage: Query only
:
CALibration<hw>:FREQuency[:MEASure]?
The command starts all adjustments which affect the frequency.
Return values:
<Measure> 0 | 1
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Usage: Query only
:CALibration<hw>:IQModulator:FULL?
The command starts adjustment of the I/Q modulator for the entire frequency range. The
I/Q modulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Return values:
<Full> 0 | 1
Example:
CAL:IQM:FULL?
starts adjustments for the I/Q modulator across the entire fre-
quency range.
Response: "0"
adjustment has been performed successfully
Usage: Query only
:CALibration<hw>:IQModulator:LOCal?
The command starts adjustment of the I/Q modulator for the currently set frequency, I/Q
swap and baseband gain. The I/Q modulator is adjusted with respect to carrier leakage,
I/Q imbalance and quadrature.
Return values:
<Local> 0 | 1
CALibration Subsystem