Instrument Function
R&S
®
SMBV100A
188Operating Manual 1407.6062.32 ─ 08
Do not save the list at this point, because the frequency entries are lost as long as there
are no entries for the level column also. In the following these entries are automatically
acquired by the connected power sensor.
All level correction values for the given frequency values are measured using the Power
Sensor and automatically filled in the selected list after the "Execute" button is pressed.
The list is automatically stored and recalled again after filling.
Fill User Correction Data with Sensor Settings
The "Fill with Sensor" button of the "Edit User Correction Data" menu opens the associ-
ated menu.
The menu indicates the relevant generator settings.
Fill User Correction Data with Sensor
All settings are read-only, except the "Sensor". In case more than one sensor is connec-
ted to the instrument, the sensor can be selected in the "Sensor" field of menu.
The "Execute" button is only enabled if a sensor is detected and the user correction list
contains at least one frequency value.
SCPI command:
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
on page 618
5.4.11 Reverse Power Protection
The reverse power protection is tripped when the power of the external signal becomes
too high. A relay opens and interrupts the internal connection to the RF output. This
condition is indicated in the display header by the "OVERLOAD" status message.
Overload
If an "Overload" status message is indicated in the display header, reset the overload
protection by pressing the RF ON/OFF key.
RF Signal and Analog Modulations - RF Block