Remote Control Commands
R&S
®
SMBV100A
458Operating Manual 1407.6062.32 ─ 08
Example:
CAL:IQM:LOC?
starts adjustment for the I/Q modulator for the currently set fre-
quency.
Response: 0
adjustment has been performed successfully
Usage: Query only
:CALibration<hw>:LEVel:ALCTable[:MEASure]?
Performs ALC adjustments according to the predefined values in the ALC table.
Return values:
<Measure>
Example:
CAL:LEV:ALCT:MEAS?
starts adjustments according to the ALC table.
Respone: "0"
adjustment has been performed successfully.
Usage: Query only
:CALibration<hw>:LEVel[:MEASure]?
The command starts all adjustments which affect the level.
Return values:
<Measure> 0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
:CALibration<hw>:LEVel:EXTern:DATA <Data>
Selects the data used for of external level correction.
By default the instrument uses correction data obtained in the factory before delivery. In
addition, customer data can be used for external level correction. The customer data is
obtained using a R&S NRP power sensor. External level correction is a protected function
(see service manual, chapter 2, "Adjustment").
Parameters:
<Data> FACT | CUST
*RST: FACTory
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external level
correction.