Remote Control Commands
R&S
®
SMBV100A
623Operating Manual 1407.6062.32 ─ 08
Example:
CORR:CSET '/var/smbv/Lists/ucor/ucor1'
selects the table ucor1.
CORR ON
activates level correction. Correction is performed using the table
ucor1.
Usage: Event
[:SOURce<hw>]:CORRection[:STATe] <State>
The command activates/deactivates level correction. Level correction is performed using
the table which has been selected with the command CORRection:CSET:SELect.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
SOUR:CORR:CSET '/var/smbv/lists/ucor/ucor1'
selects the table ucor1.
SOUR:CORR ON
activates user correction.
[:SOURce<hw>]:CORRection:VALue?
The command requests the current value for user correction.
Return values:
<Value> float
Example:
CORR:VAL?
queries the value currently used for level correction.
Response: -3
the correction value is - 3 dB.
Usage: Query only
7.14.6 SOURce:FM Subsystem
The FM subsystem contains the commands for checking the frequency modulation.
The settings for the internal modulation source (LF generator) are made in the
SOURce:LFOutput subsystem /see chapter 7.14.11, "SOURce:LFOutput Subsys-
tem", on page 647).
For information about the required options, see chapter 5.4.12.3, "Frequency Modulation
(FM)", on page 193.
[:SOURce<hw>]:FM[:DEViation]......................................................................................624
[:SOURce<hw>]:FM:EXTernal:COUPling..........................................................................624
[:SOURce<hw>]:FM:MODE.............................................................................................624
[:SOURce<hw>]:FM:SENSitivity......................................................................................625
[:SOURce<hw>]:FM:SOURce..........................................................................................625
SOURce Subsystem