Remote Control Commands
R&S
®
SMBV100A
455Operating Manual 1407.6062.32 ─ 08
7.3.3 Creating an ASCII C-List Manually
When you create the C-list in the editor, as shown in the figure, you get an ASCII file with
the extension *.dm_iqc (see the example C-List ASCII File).
Fig. 7-2: Representation in the C-List Editor
Example: C-List ASCII File
C-List in ASCII format, corresponding to C-List editor figure 7-2:
{TYPE:SMU-CL}{COPYRIGHT:2003 Rohde&Schwarz SMU}{DATE:2011-07-19;14:58:44}
{HOP LIST:0:0;498:1;506:0}{CW MODE LIST:0:0;380:1}{LEVATT LIST 1:0:0;464:1}
{BURST LIST:0:0;122:1;270:0;582:1;924:0}
{MARKER LIST 4:0:0;706:1;764:0}{MARKER LIST 3:0:0;530:1;633:0}
{MARKER LIST 2:0:0;350:1;457:0}{MARKER LIST 1:0:0;108:1;160:0}
{CONTROL LENGTH:1000}
The tags to the traces in the C-List editor are:
HOP LIST
CW MODE LIST
LEVATT LIST 1
BURST LIST
MARKER LIST 4
MARKER LIST 3
MARKER LIST 2
MARKER LIST 1
CONTROL LENGTH
CONTROL LENGTH defines the periodicity of the total C-List. The remaining tags define
the individual markers or control traces in a combined state/starting time way within the
C-List period.
R&S Signal Generator Waveform and List Format

Table of Contents