EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Impairments

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
239Operating Manual 1407.6062.32 ─ 08
5.5.2.2 Impairments
I/Q impairment for specific impairment of the I/Q modulation is set in the "Impairments"
section.
State - Impairments I/Q Mod
Activates/deactivates I/Q impairment.
If activated, the settings for leakage, I/Q imbalance and quadrature offset become effec-
tive.
Whether or not I/Q impairment is activated is indicated in the function block.
Note: It is advisable to switch automatic level control to the Sample&Hold mode, as oth-
erwise level errors may occur (see chapter 5.4.9, "Automatic Level Control -
ALC", on page 178).
SCPI command:
[:SOURce<hw>]:IQ:IMPairment[:STATe] on page 639
Offset - I/Q Mod
Sets the carrier offset (in percent) of the amplitudes (scaled to the peak envelope power
(PEP) for the I and/or Q signal component.
An ideal I/Q modulator suppresses the carrier offset completely (offset = 0 percent). If an
offset value is entered for a component, a carrier offset with fixed amplitude is added to
the signal. In the diagram, all I values or Q values are offset by a fixed amplitude value
depending on the entered percentage.
I/Q Modulation - I/Q Mod Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals