Remote Control Commands
R&S
®
SMBV100A
667Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:LIST:POWer <Power>
The command fills the Level part of the selected list with data.
*RST does not affect data lists.
Parameters:
<Power> <Power#1>{, <Power#2>, ...} | block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data. When block data is transferred, 8 (4) bytes are
always interpreted as a floating-point number with double accu-
racy (see :FORMat[:DATA] on page 464).
Range: Minimum level to Maximum level
Default unit: dBm
Example:
LIST:SEL '/var/smbv/Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
LIST:POW 0dBm, 2dBm, 2dBm, 3dBm,..
specifies the level values in list3. The number of level values
must correspond to the number of frequency values. The previous
data is overwritten.
Usage: SCPI confirmed
[:SOURce<hw>]:LIST:POWer:POINts?
The command queries the length (in points) of the LEVel part of the selected list.
Return values:
<points> float
Example:
LIST:SEL '/var/smbv/Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
LIST:POW:POIN?
queries the number of levels in list3.
Response: 327
LIST2 has 327 level entries.
Usage: Query only
SCPI confirmed
[:SOURce<hw>]:LIST:RESet
The command resets the list to the starting point.
Example:
LIST:RES
resets the list to the starting point.
Usage: Event
SOURce Subsystem