EasyManua.ls Logo

Rohde & Schwarz SMBV100A - [: S Ource<Hw>]: L IST: I Ndex: S Tart; [: S Ource<Hw>]: L IST: I Ndex: S TOP

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
665Operating Manual 1407.6062.32 ─ 08
Example:
LIST:SEL '/var/smbv/Lists/Listmode/list3'
selects list3 for use in List mode.
FREQ:MODE LIST
activates List mode. List3 is processed.
LIST:MODE STEP
selects manual, step-by-step processing of the list.
LIST:IND 5
the frequency/level value pair with index 5 is executed.
TRIG:LIST:SOUR SING
selects triggering by means of the single trigger. The list is exe-
cuted once.
SOUR:LIST:TRIG:EXEC
triggers the processing of the selected list.
[:SOURce<hw>]:LIST:INDex:STARt <Start>
The command sets the start index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list. Only the values in the set index range
(:LIST:INDex:STARt ... :LIST:INDex:STOP) are processed in List mode.
Parameters:
<Start> float
Range: 0 to list length
*RST: 0
Example:
LIST:SEL '/var/smbv/Lists/Listmode/list3'
selects list3 for use in List mode.
LIST:IND:STAR 25
sets 25 as start index of the index range.
LIST:IND:STOP 49
sets 49 as stop index of the index range.
FREQ:MODE LIST
activates List mode. The frequency/level value pairs from index 25
to index 49 in list3 are processed. All other entries of the list are
ignored.
[:SOURce<hw>]:LIST:INDex:STOP <Stop>
The command sets the stop index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list. Only the values in the set index range
(:LIST:INDex:STARt ... :LIST:INDex:STOP) are processed in List mode.
Parameters:
<Stop> float
Range: 0 to list length
*RST: 0