EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 655

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
638Operating Manual 1407.6062.32 ─ 08
Example:
IQ:CRES 10
specifies a crest factor of 10 dB for the external analog signal.
[:SOURce<hw>]:IQ:IMPairment:IQRatio[:MAGNitude] <Magnitude>
This command sets the ratio of I modulation to Q modulation (amplification “imbalance”).
Parameters:
<Magnitude> float
Range: -1.0dB to +1.0dB
Increment: 0.005dB
*RST: 0
Default unit: 0dB
Example:
IQ:IMP:IQR 0.5 dB
sets the imbalance.
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I <I>
This command sets the carrier offset for the I-channel of the I/Q modulation.
Parameters:
<I> float
Range: -10 to 10
Increment: 0.02PCT
*RST: 0
Default unit: PCT
Example:
IQ:IMP:LEAK:I 3 PCT
sets the leakage for the I-channel to 3 percent.
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q <Q>
This command sets the carrier leakage amplitude for the Q-channel of the I/Q modulation.
Parameters:
<Q> float
Range: -10 to 10
Increment: 0.02PCT
*RST: 0PCT
Default unit: PCT
Example:
IQ:IMP:LEAK:Q 3 PCT
sets the leakage for the Q-channel to 3 percent.
[:SOURce<hw>]:IQ:IMPairment:QUADrature:[ANGLe] <Angle>
This command sets the quadrature offset for the I/Q modulation.
SOURce Subsystem