Remote Control Commands
R&S
®
SMBV100A
641Operating Manual 1407.6062.32 ─ 08
Example:
IQ:OUTP:BIAS:OPT:STAT ON
activates optimization for RF output.
[:SOURce<hw>]:IQ:OUTPut:LEVel <Level>
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP <Pep>
This command sets the off-load voltage.
Single ended Output:
Value range: 1 mV ... 1.5 V EMF.
Differential Output:
Value range: 2 mV ... 3 V EMF.
Parameters:
<Pep> float
Range: 0.02 to 2
Increment: 0.001
*RST: 1
Default unit: V
Example:
IQ:OUTP:POW:PEP 2 V
sets a off-load level of 2 V
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP:RESult?
Displays the resulting output off-load voltage of both signal components at the rear panel.
Return values:
<Result> float
Range: 0.02 to 4
Increment: 0.001
*RST: 1
Usage: Query only
[:SOURce<hw>]:IQ:OUTPut:MODE <Mode>
This command selects the mode for setting the outputs.
Selection FIXed locks all IQ:OUTPut-commands except for IQ:OUTPut:TYPE and sets
them to predefined values ( *RST).
SOURce Subsystem