Instrument Function
R&S
®
SMBV100A
217Operating Manual 1407.6062.32 ─ 08
Note: Activating the RF level sweep mode automatically disables other sweep modes,
and the list mode.
SCPI command:
[:SOURce<hw>]:POWer:MODE on page 680
Mode - Level Sweep
Selects the level sweep instrument operating mode and the sweep mode.
"Auto"
Sets an automatically repeated sweep cycle. If a different sweep mode
was activated prior to the "Auto" mode, the cycle continues from the
current sweep setting.
The "Reset Sweep" button resets the sweep to the start level.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR AUTO
SOUR:POW:MODE SWE
"Single"
Sets a single sweep cycle. The sweep is triggered by the "Execute Sin-
gle Sweep" button.
If a different sweep mode was activated prior to the "Single" mode, the
current sweep is stopped. The single sweep always starts at the start
level.
The "Reset Sweep" button resets the sweep to the start level.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE