Remote Control Commands
R&S
®
SMBV100A
559Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:DELay <Delay>
The command specifies the trigger delay (expressed as a number of symbols) for external
triggering.
Parameters:
<Delay> float
Range: 0 to 65535
*RST: 0
Default unit: symbol
Example:
BB:DM:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector.
BB:DM:TRIG:DEL 50
sets a delay of 50 symbols for the trigger.
[:
SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit <Inhibit>
The command specifies the number of symbols by which a restart is to be inhibited fol-
lowing a trigger event. This command applies only in the case of external triggering.
Parameters:
<Inhibit> float
Range: 0 to 67108863
*RST: 0
Default unit: symbol
Example:
BB:DM:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector.
BB:DM:TRIG:INH 200
sets a restart inhibit for 200 symbols following a trigger event.
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence <Sequence>
The command selects the trigger mode.
SOURce Subsystem