Remote Control Commands
R&S
®
SMBV100A
588Operating Manual 1407.6062.32 ─ 08
Parameters:
<Q> float
Range: -10 to 10
Increment: 0.01
*RST: 0
Default unit: PCT
Example:
BB:IMP:LEAK:Q 3 PCT
sets the leakage for the Q-component to 3 percent.
[:SOURce<hw>]:BB:IMPairment:OPTimization:MODE <Mode>
This command sets the optimization mode.
Parameters:
<Mode> FAST | QHIGh
FAST
Optimization is reached by compensation for I/Q skew.
QHIGh
Optimization is reached by compensation for I/Q skew and fre-
quency response correction.
*RST: FAST
Example:
BB:IMP:OPT:MODE FAST
sets the optimization mode Fast.
[:SOURce<hw>]:BB:IMPairment:OPTimization:STATe <State>
This command activates and deactivates internal compensation of signal distortions by
the I/Q modulator.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
BB:IMP:OPT:STAT ON
activates internal compensation of signal distortions.
[:SOURce<hw>]:BB:IMPairment:QUADrature[:ANGLe] <Angle>
This command sets the quadrature offset for the digital I/Q signal.
Parameters:
<Angle> float
Range: -10 to 10
Increment: 0.01
*RST: 0
Default unit: DEG