EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 618

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
601Operating Manual 1407.6062.32 ─ 08
Parameters:
<Inhibit> integer
Range: 0 to 67108863
Increment: 1
*RST: 0
Example:
BB:MCCW:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector.
BB:MCCW:TRIG:INH 200
sets a restart inhibit for 200 samples following a trigger event.
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe?
The command queries the current status of signal generation for all trigger modes with
Multi Carrier CW on.
Return values:
<RMode> STOP | RUN
RUN
The signal is generated. A trigger event occurred in the triggered
mode.
STOP
The signal is not generated. A trigger event did not occur in the
triggered modes, or signal generation was stopped by the com-
mand BB:MCCW:TRIG:ARM:EXECute (armed trigger modes
only).
Example:
SOURBB:MCCW:TRIG:SOUR EXT
sets external triggering via the TRIGGER 1 connector.
SOURBB:MCCW:TRIG:MODE ARET
selects the Armed_Retrigger mode
SOURBB:MCCW:TRIG:RMOD?
queries the current status of signal generation.
Response: RUN
the signal is generated, an external trigger was executed.
Usage: Query only
[:SOURce<hw>]:BB:MCCW:TRIGger:SLENgth <SLength>
The command defines the length of the signal sequence to be output in the "Single" trigger
mode. The input is to be expressed in samples. It is then possible to output deliberately
just part of the waveform, an exact sequence of the waveform, or a defined number of
repetitions of the waveform.
Parameters:
<SLength> integer
Range: 1 to 1000
Increment: 1
*RST: 1
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals