EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 648

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
631Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:FREQuency:OFFSet <Offset>
The command sets the frequency offset of a downstream instrument, e.g. a mixer. If a
frequency offset is entered, the frequency entered with SOURce:FREQuency:... no
longer corresponds to the RF output frequency. The following correlation applies:
SOURce:FREQuency:... = RF output frequency + SOURce:FREQuency:OFFSet.
Entering an offset does not change the RF output frequency, but rather the query value
of SOURce:FREQuency:....
Parameters:
<Offset> float
Range: -67GHz to 67GHz
Increment: 0.01Hz
*RST: 0Hz
Default unit: Hz
Example:
FREQ:OFFS 500kHz
sets the frequency offset to 500 kHz.
Usage: SCPI confirmed
[:SOURce<hw>]:FREQuency:SPAN <Span>
This command specifies the span for the sweep. This setting in combination with the
center frequency setting ([SOUR]:FREQ:CENT) defines the sweep range.
This command is linked to the commands [SOUR]:FREQ:STAR and
[:SOUR]:FREQ:STOP, i.e. changing these values causes the SPAN value to change,
and vice versa:
SPAN = (STOP - STARt)
Negative values for SPAN are permitted; STARt > STOP then applies.
Parameters:
<Span> float
Range: 0 to RFmax
Increment: 0.01Hz
*RST: 400MHz
Example:
FREQ:CENT 400 MHz
sets the center frequency of the frequency sweep to 400 MHz.
FREQ:SPAN 200 MHz
sets a span of 200 MHz. This sets the sweep range to 300 MHz
to 500 MHz.
Usage: SCPI confirmed
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A