Remote Control Commands
R&S
®
SMBV100A
682Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:POWer:SPC:PEAK <PowCntrlPeak>
Activates power control by means of the peak power values, provided the power sensor
supports this function.
Parameters:
<PowCntrlPeak> 0 | 1 | OFF | ON
*RST: 0
Example:
POW:SPC:PEAK ON
uses the measured peak power for power control.
[:
SOURce<hw>]:POWer:SPC:SELect <PowCntrlSelect>
Defines the currently selected sensor to be used for power control.
Parameters:
<PowCntrlSelect> SENS1 | SENS2 | SENS3 | SENS4
*RST: SENS1
Example:
POW:SPC:SEL SENS2
selects the sensor connected to a second USB interface for power
control.
[:SOURce<hw>]:POWer:SPC:STATe <PowCntrlState>
Activates power control using the selected sensor. The control loop periodically adjusts
the generator output. After switching off, the running loop is completed.
Parameters:
<PowCntrlState> 0 | 1 | OFF | ON
*RST: 0
Example:
POW:SPC:STAT ON
activates power control.
[:SOURce<hw>]:POWer:SPC:TARGet <PowCntrlTarget>
Sets the nominal level expected at the input of the sensor. To define the unit of the power
value, use command SENSe<ch>:UNIT[:POWer] on page 493.
Parameters:
<PowCntrlTarget> float
Range: -50 to 30
Increment: 0.01
*RST: -10
Example:
SENS:UNIT dBm
selects unit dBm for setting the target level value.
POW:SPC:TARG -10
sets -10 dBm target level.