Remote Control Commands
R&S
®
SMBV100A
712Operating Manual 1407.6062.32 ─ 08
:SYSTem:DISPlay:UPDate <Update>
The command switches the update of the display on/off. A switchover from remote control
to manual control always sets the status of the update of the display to ON.
Parameters:
<Update> 0 | 1 | OFF | ON
*RST: ON
Example:
SYST:DISP:UPD OFF
switches update of displayed parameter values off.
:
SYSTem:DLOCk <Dlock>
This command (Display LOCk) disables the manual operation via the display, or enables
it again (OFF).
The command disables also the front panel keyboard of the instrument including the
LOCAL key.
Parameters:
<Dlock> 0 | 1 | OFF | ON
*RST: 0
Example:
SYST:DLOC ON
activates the display lock. The instrument cannot be operated via
the display until it has been enabled with SYST:DLOC OFF.
:SYSTem:FILes:TEMPorary:DELete
The command deletes the temporary files from the directory /var/smbv/unicod.
The content of this directory are temporary files such as the wavefiles (*.wv) of all ARB
standards. Deleting the temporary files maintains sufficiant amount of free disc space on
the internal flash memory.
Another way to achieve more free disc space on the internal flash memory is to execute
the SYST:FPReset command.
Example:
SYST:FIL:TEMP:DEL
The content of the folder /var/smbv/unicod is deleted.
Usage: Event
SCPI confirmed
SYSTem:ERRor:ALL?
The command queries all entries in the error queue and then deletes them.
SYSTem Subsystem