Remote Control Commands
R&S
®
SMBV100A
715Operating Manual 1407.6062.32 ─ 08
Parameters:
<Identification> AUTO | USER
*RST: AUTO
Example:
SYST:IDEN USER
selects the user defined identification string.
SYST:IRES "Test Device"
defines the identification string 'test device'
*IDN?
Response: 'test device'
SYST:LANG 'HP8657'
selects command set of the HP generator. The R&S SMBV's com-
mand set is disabled.
To return to the SCPI command set of the R&S SMBV, use the HP
command EX.
:SYSTem:IRESponse <Iresponse>
The command defines the identification string for selection user defined (SYST:IDEN
USER).
Note: While working in a emulation mode, the instrument's specific command set is dis-
abled, i.e. the SCPI command SYST:IRES will be discarded.
Parameters:
<Iresponse> string
Example:
SYST:IDEN USER
selects an user-defined identification
SYST:IRES "Test Device"
defines the identification string 'test device'
*IDN?
Response: 'test device'
:SYSTem:KLOCk <Klock>
This command (Keyboard LOCk) disables the front panel keyboard of the instrument
including the LOCAL key, or enables it again (OFF).
Parameters:
<Klock> 0 | 1 | OFF | ON
*RST: OFF
Example:
SYST:KLOC ON
activates the keyboard lock. The keyboard cannot be operated
again until it has been enabled with SYST:KLOC OFF.
:SYSTem:LANGuage <Language>
Sets the remote control command set.