097-55501-02 Issue 7 – January, 2005 TimeHub 5500 TL1 Reference Guide 49
Commands
Command Description
Command Examples:
ED-SYNC::S1-OUTA-3:SC001:::SQUELCH=Y;
The command in this example turns off output channel 3 of Output card A in shelf 1.
ED-SYNC::CLK-5:SC001:::PRIORITY=1,ENABLE=Y;
The command in this example sets the priority of input channel 5 and enables the
input channel.
ED-SYNC::CLK-0&&4:SC001:::ENABLE=Y;
The command in this example enables input channels 0 through 4, inclusive.
ED-SYNC::CLK-3:SC001:::TDEV-1S=THR-100;
The command in this example sets the threshold for the 1 second TDEV
measurement on input channel 3 to 100 nanoseconds.
ED-SYNC::CLK-3:SC001:::LMRTIE-15M=THR-500&CRI-Y&RPT-Y&PM-Y;
The command in this example specifies the LMRTIE-15M threshold to be
500 nanoseconds for input channel 3 (THR-500). It also disqualifies input channel 3
as an input reference (CRI-Y). In addition, if the threshold is exceeded the system
displays an Automatic Message (RPT-Y) and displays the measurement every 15
minutes (PM-Y).
Edit User Security
This command changes the user name, access level, or password. The command
format is:
ED-USER-SECU:[<tid>]:<aid>:<ctag>::<newuid>,<newpid>,,<uap>;
<measurement> FFOF — Fractional Frequency Offset
LMRTIE Latest Maximum Relative Time Interval
Error
MRTIE Maximum Relative Time Interval Error
SPRD Frequency Spread
TDEV Time Deviation
TIE Time Interval Error
<value> See Chapter 3, Keywords.
Parameter Value Variables Description