Sysmex XP-300 12-13
Cleaning and Maintenance
Revised July 2012
(8) Remove the rotary valve.
(9) Clean the rotary valve using distilled water or 1:10 dilution of
CELLCLEAN detergent. After cleaned with CELLCLEAN,
always clean it with distilled water.
(10) Clean the contact surfaces of the fixed and rotary valves using a
gauze moistened with distilled water. By using CELLCLEAN
together with distilled water, stuck objects, dirt, etc. can be
removed easily.
(11) Make sure the valve contact surfaces are completely free from
dirt or dust.
Rotary valve
Fixed valve
Caution!
When removing each valve component, reagent could
leak from the tube. If it does, wipe it clean using cloth.
If left as it is, it can cause current leakage or electric
shocks.
Note:
The valve components are in close contact with one
another. They can be easily removed when you slide each
one while twisting.
Caution!
Take care not to inflict flaws or scratches on valve
surfaces, since flaws or scratches can cause blood
leakage and correct analysis results may not be
obtained.
Do not use any other materials, like cleaning tissues
or Q-tips for cleaning which could loose parts of its
substance.
Information
Do not use any detergent other than CELLCLEAN.
Although the SRV is corrosion-resistant against
CELLCLEAN, rinse the SRV with distilled water
completely to prevent troubles to the unit or other
components.
Caution!
If the device is used with dirt or dust attaching on
valve contact surfaces, blood leakage can occur and
correct analysis results may not be obtained.