13-8 Sysmex XP-300
Troubleshooting
Revised April 2012
2. Chamber Errors
Error Message
(Error Code)
Pressure/Vac Error (119010)
Cause Pressure fell below 0.02 MPa due to the following reason(s):
0.05 MPa pressure adjustment error
Pneumatic unit pressure error
Air leakage from pressure line
Action Adjust pressure to 0.05 MPa.
Turn the main power switch OFF, then ON again. After that, adjust pressure by
referring to “Adjusting pressure to 0.05 MPa” in “12.15 Adjust pressure and
vacuum”.
Note The power needs to be turned OFF then ON again.
The status is not ready for analysis, but the processing, etc. of stored data can
be performed.
Error Message
(Error Code)
Waste not drained (141000)
Cause Breakage or clogging of waste chamber or waste line tube
Pneumatic unit pressure error
Air leakage from pressure line
Float switch malfunction
Faulty operation of solenoid valve or master valve
Action Check waste line tube.
Check the waste drain line for breakage, clogging, etc. If any abnormality is
present, clean or replace the tube. Specially check for contaminant or clogging
near the drain outlet nipple.
Troubleshooting
on the screen
After adjustment and checking, press the [Execute] button to
clear the error, and the waste chamber drain sequence is
executed.
Note Not ready until the error is resolved.
The status is not ready for analysis, but the processing, etc. of stored data can
be performed.