Sysmex XP-300 13-9
Troubleshooting
Revised April 2012
Error Message
(Error Code)
Replace CELLPACK (131000)
Cause Lack of diluent (CELLPACK)
Diluent aspiration line error
Air leakage from vacuum line
Float switch malfunction
Faulty operation of solenoid valve or master valve
Use in excess of the specified number of analyses for a container
Action Replenish diluent (CELLPACK).
If there is no diluent or the volume is insufficient, replace with a new diluent
container.
Check tube.
Check the diluent line tubing for breakage and the nipple for looseness. If any
abnormality is noted, reconnect or replace.
Troubleshooting
on the screen
Replace with a new reagent and press the [Execute] button.
The error is cleared and the reagent replacement sequence is
executed.
Refer to “12.16 Replace reagent”.
However, an error occurs again if the reagent replacement
sequence is not completed correctly.
Note Not ready until the error is resolved.
The status is not ready for analysis, but the processing, etc. of stored data can
be performed.
Error Message
(Error Code)
Replace STROMATOLYSER (133220)
Cause Lack of lyse (STROMATOLYSER-WH)
Float switch malfunction
Use in excess of the specified number of analyses for a bottle.
Action Replenish lyse (STROMATOLYSER-WH).
If there is no lyse or the volume is insufficient, replace with a new lyse
container.
Check the lyse tubing for breakage. If any abnormality is noted, reconnect or
replace.
Troubleshooting
on the screen
Replace with a new reagent and press the [Execute] button.
The error is cleared and the reagent replacement sequence is
executed.
Refer to “12.16 Replace reagent”.
However, an error occurs again if the reagent replacement
sequence is not completed correctly.
Note Not ready until the error is resolved.
The status is not ready for analysis, but the processing, etc. of stored data can
be performed.