13-10 Sysmex XP-300
Troubleshooting
Revised July 2012
Error Message
(Error Code)
CELLPACK expired (472020)
Cause Expiration date of the CELLPACK has expired.
Action Replace the CELLPACK.
Replace with a CELLPACK of which expiration date has not expired.
Troubleshooting
on the screen
Replace with a new reagent and press the [Execute] button.
The error is cleared and the reagent replacement sequence is
executed.
Refer to “12.16 Replace reagent”.
However, an error occurs again if the reagent replacement
sequence is not completed correctly.
Error Message
(Error Code)
S.LYSER expired (472034)
Cause Expiration date of the STROMATOLYSER has expired.
Action Replace the STROMATOLYSER.
Replace with a STROMATOLYSER of which expiration date has not
expired.
Troubleshooting
on the screen
Replace with a new reagent and press the [Execute] button.
The error is cleared and the reagent replacement sequence is
executed.
Refer to “12.16 Replace reagent”.
However, an error occurs again if the reagent replacement
sequence is not completed correctly.